Publication:
Dose estimation on surgeon for radionuclide vertobroplasty

Дата
2021
Авторы
Zabaryansky, Yu. G.
Kurachenko, Yu. A.
Golovin, A. A.
Journal Title
Journal ISSN
Volume Title
Издатель
Научные группы
Организационные подразделения
Выпуск журнала
Аннотация
© 2021 State Research Center, Burnasyan Federal Medical Biophysical Center of Federal Medical Biological Agency. All rights reserved.Purpose: To estimate the doses to a surgeon for a new operation - radionuclide vertebroplasty. To normalize the irradiation for the maximum number of operations in which the total equivalent dose will not exceed the values specified in the NRB-99/2009. Material and methods: The widespread Monte-Carlo code MCNP used with data libraries, as well as the ICRP-38 data on radionuclide characteristics. Results: Using the radionuclide 188Re, the most promising for this procedure, the equivalent dose in the doctor's palms does not exceed 1 mSv per operation. A surgeon can perform up to 580 operations per year without violating radiation safety standards. Conclusion: Calculation methods proved the safety of radionuclide vertebroplasty for the surgeon and service personnel.
Описание
Ключевые слова
Цитирование
Zabaryansky, Yu. G. Dose estimation on surgeon for radionuclide vertobroplasty / Zabaryansky, Yu.G., Kurachenko, Yu.A., Golovin, A.A. // Medical Radiology and Radiation Safety. - 2021. - 66. - № 2. - P. 76-77. - 10.12737/1024-6177-2021-66-2-76-77
Коллекции